皮膚ガン診断用サーミスタ特性校正に用いる真空断熱式温度校正装置の評価
皮膚ガン診断用サーミスタ特性校正に用いる真空断熱式温度校正装置の評価
カテゴリ: 論文誌(論文単位)
グループ名: 【A】基礎・材料・共通部門
発行日: 2024/04/01
タイトル(英語): Evaluation of a Vacuum Insulated Temperature Calibration Device to Calibrate Thermistor Characteristics for Skin Cancer Diagnosis
著者名: 藤田 法明(八戸工業高等専門学校),野中 崇(八戸工業高等専門学校),細川 靖(八戸工業高等専門学校),古川 琢磨(八戸工業高等専門学校),郭 福会(八戸工業高等専門学校),井関 祐也(八戸工業高等専門学校),横田 実世(八戸工業高等専門学校),岡部 孝裕(弘前大学大学院理工学研究科),圓山 重直(東北大学総務企画部)
著者名(英語): Homei Fujita (National Institute of Technology, Hachinohe College), Takashi Nonaka (National Institute of Technology, Hachinohe College), Yasushi Hosokawa (National Institute of Technology, Hachinohe College), Takuma Kogawa (National Institute of Technology, Hachinohe College), Fuhui Guo (National Institute of Technology, Hachinohe College), Yuya Iseki (National Institute of Technology, Hachinohe College), Miyo Yokota (National Institute of Technology, Hachinohe College), Takahiro Okabe (Graduate School of Science and Technology, Hirosaki University), Shigenao Maruyama (General Affairs and Planning Department, Tohoku University)
キーワード: サーミスタ,温度校正,皮膚ガン thermistor,temperature calibration,skin cancer
要約(英語): Recent methods using thermistors as an early detection of skin cancer require the higher resolution of 5 mK. Previously, the authors developed a vacuum insulated temperature device that calibrates the resistance temperature characteristics of thermistors, so that it improves the temperature resolution and accuracy of skin cancer diagnosis. In this study the operational methods of the calibration system were compared for improving further accuracy using the temperature calibration device and the oil tank calibration device. The results of the comparison indicated that the temperature calibration device is probably as accurate as the oil tank calibration device.
本誌: 電気学会論文誌A(基礎・材料・共通部門誌) Vol.144 No.4 (2024) 特集:2023年基礎・材料・共通部門大会
本誌掲載ページ: 126-131 p
原稿種別: 論文/日本語
電子版へのリンク: https://www.jstage.jst.go.jp/article/ieejfms/144/4/144_126/_article/-char/ja/