ドップラレーダを用いたポータブル型呼吸計測器の開発
ドップラレーダを用いたポータブル型呼吸計測器の開発
カテゴリ: 論文誌(論文単位)
グループ名: 【C】電子・情報・システム部門
発行日: 2023/09/01
タイトル(英語): Development of a Stand-alone Portable Respiratory Rate Monitor Using a Doppler Radar
著者名: 加藤 美夕((株)日立ハイテク 事業開発本部),金澤 建(昭和大学江東豊洲病院 こどもセンター),阿部 祥英(昭和大学江東豊洲病院 こどもセンター),孫 光鎬(電気通信大学 情報理工学研究科),松井 岳巳(東京都立大学大学院 システムデザイン研究科)
著者名(英語): Miyu Katoh (Business Development Division, Hitachi High-Tech Corporation), Takeru Kanazawa (Children's Medical Center, Showa University Koto Toyosu Hospital), Yoshifusa Abe (Children's Medical Center, Showa University Koto Toyosu Hospital), Guanghao Sun (
キーワード: ドップラレーダ,非接触,呼吸数,マイクロ波 doppler radar,non-contact,respiratory rate,microwave
要約(英語): Respiratory rate (RR) is known to be a more accurate predictor of clinical deterioration than other vital signs. However, there were few respiration measurement devices certified as medical devices that could be used in daily clinical settings. Therefore, using a bult-in microcontroller (72×57×12 mm) instead of a personal computer, we developed a portable stand-alone respiration measurement device with minimum workload in computing that can be used for non-contact measurement in 30 seconds using a Doppler radar.In this study, the problems of respiration measurement using a Doppler radar, such as miscounting of respiratory peaks were clarified, and proposed device with respiratory peaks miscount prevention algorithm achieved high accuracy RR measurement. Clinical testing was conducted on pediatric outpatients of Children's Medical Center, Showa University Koto Toyosu Hospital. The measurement accuracy of the system was confirmed to be comparable to the respiration measurement accuracy of stationary certified medical devices used in hospital, such as capnometers and the chest wall impedance method used in bedside monitors.
本誌: 電気学会論文誌C(電子・情報・システム部門誌) Vol.143 No.9 (2023) 特集:知能メカトロニクス分野と連携する知覚情報技術
本誌掲載ページ: 922-933 p
原稿種別: 論文/日本語
電子版へのリンク: https://www.jstage.jst.go.jp/article/ieejeiss/143/9/143_922/_article/-char/ja/
受取状況を読み込めませんでした